30W Medical Grade Switching Power Supplies
AHP3201 series
WWW.ASTRODYNETDI.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
Wide 80-275VAC Input, 47 to 63Hz
IEC60601-1 3rd Edition: 2MOPP
Choice of 2 Input Congurations
Single Output
Efciency Level V
Class II
Model Number Output Watts Output Voltage Output Amps Efciency Total Regulation Ripple & Noise
*AHP3201(A,B)-02 20 5VDC 4.0 80% ±5% 50mV pk-pk
*AHP3201(A,B)-03 23 6VDC 3.83 80% ±5% 60mV pk-pk
*AHP3201(A,B)-04 27 10VDC 2.7 81% ±5% 80mV pk-pk
*AHP3201(A,B)-05 30 12VDC 2.5 81% ±5% 100mV pk-pk
*AHP3201(A,B)-06 30 15VDC 2.0 83% ±5% 100mV pk-pk
*AHP3201(A,B)-07 30 18VDC 1.66 83% ±5% 100mV pk-pk
*AHP3201(A,B)-08 30 24VDC 1.25 85% ±3% 100mV pk-pk
*AHP3201(A,B)-09 30 30VDC 1.0 85% ±3% 100mV pk-pk
*AHP3201(A,B)-10 30 36VDC 0.83 86% ±3% 100mV pk-pk
*AHP3201(A,B)-11 30 48VDC 0.625 86% ±3% 100mV pk-pk
*Note:
The Asterisk
* denotes a choice of receiving this product with an IEC-320-C18, 2 Prong Input Receptacle (A) or with an IEC-320-C8, 2 Pole Input Receptacle (B).
See the Mechanical les at the end of this specication for clarity. Simply add the letter “A” or “B” after “AHP3201” in the model number (i.e. “AHP3201A-05” for C18
or “AHP3201B-05” for C8) when ordering to receive the desired version.
medical
HPU32Aseries
30WExternalMedicalGradePowerSupply
TheHPU32AseriesofAC/DCswitchingmodepowersupplies
provide30Wattsofcontinuousoutputpower.Allsuppliesare
UL94V-1mincompliant.All modelsmeetFCCPart-18classBand
CISPR-11EN55011classBemissionLimitsandaredesignedto
complywithANSI/AAMIES60601-1:2005(UL/cUL3rd Edition),EN
60601-1:2006(TUV/T-mark3rdEdition)andnewCErequirements.
Allunitsare100%burnedinandtested.
FEATURES:
APPLICATIONS:
GENERALSPECIFICATION:
*WideOperatingVoltage,80to275VAC,47to63Hz
*IEC-320-C18InputInlet
*SingleOutput
*OverVoltageandOverLoadprotectio n
*MedicalSafety3rd(IEC60601-13rdEdition)
*InputtoOutput:2MOPP
*EnergyStar2.0,EfficiencylevelV
*ClassIIsystem
*3yearwarranty
*HomeHealthcareEquipment
*PatientMonitor
*BloodPressuresystem
*Portablemedicaldevices
*ECGmachine
* :AutoRecovery
* :FreeAirConvection
* :UL94V-1
* :Doubleinsulated,ClassII
:ANSI/AAMIES60601-1:2005(UL/cUL3rdEdition),EN60601-1:2006
(TUV/T-mark3rdEdition)
ShortCircuitProtection
Cooling
FlammabilityRating
ProtectionClasses
*Safety
ElectricalCharacteristics:
Environmental:
Symbol
Characteristic
Condition
Min.
Typ. Max.
Unit
VAC
Class
VAC
VAC
Hz
W
A
A
A
A
%
%/°C
°C
RH
RH
kV
kV
kV
h
m
G
°C
%
ms
ms
s
Symbol
Characteristic
Condition
Min.
Typ. Max.
Unit
Vins
SafetyApprovalInputVoltageRange SafetyApproval&SpecificationinLabel
Vin
InputOperateVoltageRange
InputFrequency
DetailtoseeFig.1
(Derateli nearl yfrom100%loadat90VACto80%loadat80VAC)
Fi
Sinewave
Po
OutputPowerRange
SeeRatingChart
Iil
LowLineInputCurrent
FullLoad,Vin=100VAC
Iih
HighLineInputCurrent
FullLoad,Vin=240VAC
Irl
LowLineInputInrushCurrent
FullLoad,25°C,Coolstart,Vin=100VAC
Irh
HighLineInputInrushCurrent
FullLoad,25°C,Coolstart,Vin=240VAC
Efficiency
FullLoad,Vin=230VAC,DetailtoseeRatingChart
△Voi
LineRegulation
FullLoad,Vin=100~120VACor200~240VAC
OLP
OverLoadProtection
Recoversautomaticallyafterfaultconditionisremoved
ttr
TimeofTransientResponse
FullLoad,Vin=110VAC
thu
Hold-UpTime
FullLoad,Vin=100VAC
ts
Start-uptime
FullLoad,Vin=100~240VAC
Tc
TemperatureCoefficient
AllCondition
HV
DielectricWithstandingVoltage(P-S)
PrimarytoSecondary,limitcurrent<10mA
EMI
To
Ts
Ho
Hs
Vsg
ESDa
ESDc
MTBF
ELEV
VBR
EMCEmission
CompliancetoEN55011(CISPR11),EN61000-3-2,-3
100
80
47 63
30
0.6
0.4
28
67
SeeRatingChart
110
12
4000
-10
0
0
2
8
6
100k
5
-40
B
150
70
85
95%
95%
3000
4
20.3
±0.04
1
240
275
OperatingTemperature
StorageTemperature
OperatingHumidity
StorageHumidity
SurgeVoltage
ElectroStaticDischarge
ElectroStaticDischarge
MeanTimeBetweenFailure
OperatingAltitude(Elevation)
Vibration
DetailtoseeFig.2
(Derateli nearl yfrom100%loadat40°Cto50%loadat70°C)
10~95%RH
non-condensing
SeeRatingChart
AllCondition
AirDischarge,IEC61000-4-2
ContactDischarge,IEC61000-4-2
OperatingTemperatureat25°C,CalculatedperMIL-HDBK-217F
Allcondition
10~500Hz,10min./1cycle,60min.eachalongX,Y,Zaxes
RoHS
2011/65/EU
2
CB
S
E
P
RT
EN 60601-1
IEC 60601-1
APPROVALS:
%
OVP
112
132
OverVoltageProtection
Input Option “A” Input Option “B”
medical
HPU63Bseries
63WExternalMedicalGradePowerSupply
TheHPU63BseriesofAC/DCswitchingmodepowersupplies
provide63Wattsofcontinuousoutputpower.Allsuppliesare
UL94V-1mincompliant.All modelsmeetFCCPart-18classBand
CISPR-11EN55011classBemissionLimitsandaredesignedto
complywithANSI/AAMIES60601-1:2005(UL/cUL3rd Edition),EN
60601-1:2006(TUV/T-mark3rdEdition)andnewCErequirements.
Allunit sare100%burnedinandtested.
FEATURES:
APPLICATIONS:
GENERALSPECIFICATION:
*WideOpe ratingVoltage,80to275VAC,47to63Hz
*IEC-320-C8InputInlet
*SingleOutput
*OverVoltageandOverLoadprotection
*MedicalSafety3rd(IEC60601-13rdEdition)
*EnergyStar2.0,EfficiencylevelV
*ClassIIsystem
*3yearwa rranty
*HomeHealthcareEquipment
*PatientMonitor
*BloodPressuresystem
*Portablemedicaldevices
*ECGmachine
* :AutoRecovery
* :FreeAirConvection
* :UL94V-1
* :Doublei nsulated,ClassII
:ANSI/AAMIES60601-1:2005(UL/cUL3rdEdition),EN60601-1:2006
(TUV/T-mark3rdEdition)
ShortCircuitProtection
Cooling
FlammabilityRating
ProtectionClasses
*Safety
ElectricalCharacteristics:
Environmental:
Symbol
Characteristic
Condition
Min.
Typ. Max.
Unit
VAC
Class
VAC
VAC
Hz
W
A
A
A
A
%
%/°C
°C
RH
RH
kV
kV
kV
h
m
G
°C
%
ms
ms
s
Symbol
Characteristic
Condition
Min.
Typ. Max.
Unit
Vins
SafetyApprovalInputVoltageRange SafetyApproval&SpecificationinLabel
Vin
InputOperateVoltageRange
InputFrequency
DetailtoseeFig.1
(Deratelinearlyfrom100%loadat90VACto80%loa dat80VAC)
Fi
Sinewave
Po
OutputPowerRange
SeeRatingChart
Iil
LowLineInputCurrent
FullLoad,Vin=100VAC
Iih
HighLineInputCurrent
FullLoad,Vin=240VAC
Irl
LowLineInputInrushCurrent
FullLoad,25°C,Coolstart,Vin=100VAC
Irh
HighLineInputInrushCurrent
FullLoad,25°C,Coolstart,Vin=240VAC
Efficiency
FullLoad,Vin=230VAC,DetailtoseeRatingChart
△Voi
LineRegulation
FullLoad,Vin=100~120VACor200~240VAC
OLP
OverLoadProtection
Recoversautomaticallyafterfaultconditionisremoved
ttr
TimeofTransientResponse
FullLoad,Vin=110VAC
thu
Hold-UpTime
FullLoad,Vin=100VAC
ts
Start-uptime
FullLoad,Vin=100~240VAC
Tc
TemperatureCoefficient
AllCondition
HV
DielectricWithstandingVoltage(P-S)
PrimarytoSecondary,limitcurrent<10mA
EMI
To
Ts
Ho
Hs
Vsg
ESDa
ESDc
MTBF
ELEV
VBR
EMCEmission
CompliancetoEN55011(CISPR11),EN61000-3-2,-3
100
80
47 63
63
1.62
0.72
37
74
SeeRatingChart
110
12
0.3
4000
-10
0
0
2
8
6
100k
5
-40
B
150
70
85
95%
95%
3000
4
2
±0.04
1
240
275
OperatingTemperature
StorageTemperature
OperatingHumidity
StorageHumidity
SurgeVoltage
ElectroStaticDischarge
ElectroStaticDischarge
MeanTimeBetweenFailure
OperatingAltitude(Elevation)
Vibration
DetailtoseeFig.2
(Deratelinearlyfrom100%loadat40°Cto50%loadat70°C)
10~95%RH
non-condensing
SeeRatingChart
AllCondition
AirDischarge,IEC61000-4-2
ContactDischarge,IEC61000-4-2
OperatingTemperatureat25°C,CalculatedperMIL-HDBK-217F
Allcondition
10~500Hz,10min./1cycle,60min.eachalongX,Y,Zaxes
RoHS
2011/65/EU
2
CB
EN 60601-1
IEC 60601-1
S
E
P
APPROVALS:
%
OVP
112
132
OverVoltageProtection
30W Medical Grade Switching Power Supplies
AHP3201 series
WWW.ASTRODYNETDI.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
INPUT SPECIFICATIONS
Input Voltage Range 80-275 VAC (100-240 Nominal)
Frequency Range 47-63 Hz
Input Current Low Line (100VAC) 0.6A max.
High Line (240VAC) 0.4A max.
Inrush Current, typ (Cold Start) Low Line (100VAC) 28A max.
High Line (240VAC) 67A max.
No Load Power Consumption 0.3 Watts
OUTPUT SPECIFICATIONS
Voltage and Current See Selection Chart
Output Power Range 0 to FL Rating
Total Regulation See Selection Chart
Line Regulation (FL) (Note 3) ±1%, max.
Over Load Protection* 110-150%, Auto-Recovery
Over Voltage Protection* 112-132%
Short Circuit Protection Auto Recovery
Transient Response 110Vin, FL 4mS
Temperature Coefcient ±0.04%/°C, typ
Ripple/Noise (Note 1) See Selection Chart
Hold Up Time (Note 2) 100~240Vin, FL 12mS min.
Start Up Time 100~240Vin, FL 0.3S typ., 2S max.
GENERAL SPECIFICATIONS
Input-Out Isolation 4000 VAC, Limit Current <10mA
Input-GND 2800 VDC
Isolation Resistance 50MΩ
Efciency (FL, 230Vin) See Selection Chart
Safety ANSI/AAMI ES 60601-1:2005
(UL/cUL 3rd Edition),
EN 60601-1:2006
(TUV/T-mark 3rd Edition)
Class Class II
PHYSICAL SPECIFICATIONS
Size 4.21” x 1.85” x 1.17”
(107 x 47 x 29.6mm)
Construction Desktop Style
Weight 9.87 oz (280g)
Flammability Rating UL94V-1
ENVIRONMENTAL SPECIFICATIONS
Oper. Temperature -10 to +40°C (FL) *
50% Load @ 70°C
(See Derate Curve Last Page)
Cooling Free Air Convection
Storage Temperature -40 to +85°C, 10~95% RH *
Operating Humidity 0% to 95%, non-cond *
EMC Compliance to EN55011 (CISPR11),
EN61000-3-2, -3 Class B
MTBF (MIL-HDBK-217F) 100,000 Hrs
Surge Voltage 2kV
ESD - Air IEC61000-4-2, 8kV
ESD - Contact IEC61000-4-2, 6kV
Operating Altitude (Elevation) 3000M
Vibration 5G, 10 ~ 500Hz, 10min./1cycle,
60min. each along X, Y, Z axes
NOTES
1. Ripple & noise is measured by using 20MHz bandwidth limited
oscilloscope and terminated each output with a 0.47uF capacitor at
rated load and nominal line.
2. Hold up time is measured from the end of the last charging pulse to
the time which the main output drops down to low limit of main output at
rated load and nominal line.
3. Full Load, Vin=100~120VAC or 200~240VAC
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
All specications are typical at nominal input, full load, and 25°C unless other-
wise noted
30W Medical Grade Switching Power Supplies
AHP3201 series
WWW.ASTRODYNETDI.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
MECHANICAL DIMENSIONS - AHP3101B
DC OUTPUT CONNECTIONS
Coaxial Plug:
+Vout-Vout
5.5mm outer diameter
2.1mm inner diameter, female
11mm length, center “+”
AHP3201x-02~04 Output Cord Length 48”, 16AWG
AHP3201x-05~11 Output Cord Length 72”, 18AWG
If modied, the regulation and efciency will be affected
Dimensions are In mm [inches].
medical
HPU32Aseries
SPECIFICATIONNOTE:
1.Outputcanprovideuptopeakloadwhenthepowersupplystartsup.
Continuousstayinginmorethanratedloadisnotallowed.
2.Atfactory,in60%ratedloadcondition,eachoutputischeckedtobe
withinvoltageaccuracy.
3.Lineregulationisdefinedbychanging±10%ofinputvoltagefrom
nominallineatratedload.
4.Loadregulationisdefinedbychanging±40%ofmeasuredoutputload
from60%ratedload.
5.Ripple&noiseismeasuredbyusing20MHzbandwidthlimited
oscilloscopeandterminatedeachoutputwitha0.47uFcapacitorat
ratedloadandnominalline.
6.Holduptimeismeasuredfromtheendofthelastchargingpulsetothe
timewhichthemainoutputdropsdowntolowlimitofmainoutputat
ratedloadandnominalline.
7.Efficiencyismeasuredatratedload,andnominalline.
8.ThespecificsfortestingtheenergyefficiencyofthisSeriesareoutlined
inaseparatedocumenttitled"TestMethodforCalculatingtheEnergy
EfficiencyofSingle-VoltageInterchangeableAC-DCandAC-ACPower
Supplies(August11,2004),"whichisavailableontheENERGYSTAR
Website.
MECHANICALDIMENSIONS: (UNIT:m m)
OUTPUTCABLERECOMMEND:
PACKING:
1.Selectedoutputconnectorsandwire,pleaserefertoAppendix.
2.HPU32A-102~104arerequiredtouseAWG#16 4FToutputcable.
3.HPU32A-105~111arerequiredtouseAWG#18/6FToutputcable.
4.Theregulationandefficiencywillbechangedbymodifiedoutputcable.
1.Netweight:265~280gapprox.
2.Optionaloutputconnectorsavailablecontactsalesfordetails.
/
RatingChart:
INPUTVOLTAGE(VAC)
LOAD(%)
80
50
60
70
80
90
100
110
100 120 140 160 180 200 220 240 260 280
(FIG.1)INPUTVOLTAGEDERATINGCURVE
MODELNO.
VoltageRange
min
(VDC) (VDC) (A) (A) (W) (W) (ms)(mVp-p) (%) (%)
min
max max
OutputCurrent
(Basedontheoutputvolt.)
Maximum
OutputPower
Ripple&N oise
TotalRegulation
Typ.Efficiency
NoLoad
Consumption
Hold-UpTime
ProtectionMode
HPU32A-102 5.0 5.99 4.003.33 20 50 80 0.3 12 OLP
OLP
OLP
OLP
OLP
OLP
OLP
12
12
12
12
12
12
0.3
0.3
0.3
0.3
0.3
0.3
80
81
81
83
83
85
±5
±5
±5
±5
±5
±5
±3
60
80
100
100
100
100
23
27
30
30
30
30
3.832.87
3.382.45
2.742.30
2.301.88
1.881.43
1.431.11
8.0
11.0
13.0
6.0
8.0
11.0
13.0
16.0
16.0
21.0
21.0
27.0
HPU32A-103
HPU32A-104
HPU32A-105
HPU32A-106
HPU32A-107
HPU32A-108
OLP120.385±3100
30
1.110.9127.0 33.0HPU32A-109
OLP120.386±3100
30
0.910.7533.0 40.0HPU32A-110
OLP120.386±3100
30
0.750.6340.0 48.0HPU32A-111
30WExternalMedicalGradePowerSupply
TEMPERATURE(°C)
-10
0
10
0
10
20
30
40
50
60
70
80
90
100
110
20 30 40 50 60 70 80
(FIG.2)TEMPERATUREDERATINGCURVE